GS 04L57B01-01EN
2
<<Contents>> <<Index>>
All Rights Reserved. Copyright © 2018, Yokogawa Electric Corporation
2018.6.12-00
• Measuring range/accuracy:* See the table below.
*
Performance at standard operating conditions: 23 ± 2°C, 55 ±10%RH, normal operating conditions for other parameters.
Reference junction temperature compensation accuracy is not included for thermocouples.
No vibrations or other hindrances to performance.
Input type
Range
setting
Physical range
Measuring accuracy
6
Resolution
Standard mode
Battery-save mode
Thermocouple
K
1
-200.0 to
1370.0°C ± (0.15%of rdg + 0.7°C)
Except -200.0 to -100.0°C:
±(0.15% of rdg + 1.0°C)
±(0.2%of rdg + 3.5°C)
Except -200.0 to -100.0°C:
±(0.2% of rdg + 6.0°C)
0.1°C
E
1
-200.0 to
800.0°C
± (0.15%of rdg + 0.5°C)
Except -200.0 to -100.0°C:
±(0.15% of rdg + 0.7°C)
± (0.2%of rdg + 2.5°C)
Except -200.0 to -100.0°C:
±(2% of rdg + 5.0°C)
J
1
-200.0 to
1100.0°C
T
1
-200.0 to
400.0°C
R
1
0.0 to
1760.0°C ±(0.15% of rdg + 1.0°C)
However,
For R, S
0.0 to 100.0°C: ±3.7°C
100.0 to 300.0°C: ±1.5°C
For B
400.0 to 600.0°C: ±2.0°C
Accuracy not guaranteed
for temperatures less than
400.0°C
±(0.2% of rdg + 4.0°C)
However,
For R, S
0.0 to 100.0°C: ±10.0°C
100.0 to 300.0°C: ±5.0°C
For B
400.0 to 600.0°C: ±7.0°C
Accuracy not guaranteed
for temperatures less than
400.0°C
S
1
0.0 to
1760.0°C
B
1
0.0 to
1820.0°C
N
1
-270.0 to
1300.0°C ± (0.15% of rdg + 0.7°C)
However,
-200.0 to 0.0°C: ±(0.35% of rdg
+ 0.7°C)
Accuracy not guaranteed
for temperatures less than
-200.0°C
±(0.3%of rdg + 3.5°C)
However,
-200.0 to 0.0°C: ±(0.7% of rdg
+ 3.5°C)
Accuracy not guaranteed
for temperatures less than
-200.0°C
WRe3-25
2
0.0 to
2400.0°C ±(0.2% of rdg + 2.5°C)
Except 0.0 to 200.0°C: ±4.0°C
±(0.3% of rdg + 10.0°C)
Except 0.0 to 200.0°C: ±18.0°C
RTD
Pt100
3
-200.0 to
600.0°C
± (0.15% of rdg + 0.3°C)
±(0.3% of rdg +1.5°C)
JPt100
4
-200.0 to
550.0°C
DC voltage
20 mV
-20.000 to 20.000 mV
± (0.05% of rdg + 12 digits)
± (0.1% of rdg + 40 digits)
1 μV
60 mV
-60.00
to
60.00 mV ± (0.05% of rdg + 3 digits)
± (0.1% of rdg + 15 digits)
10 μV
200 mV
-200.00 to 200.00 mV
2 V
-2.0000 to 2.0000 V
± (0.05% of rdg + 12 digits)
± (0.1% of rdg + 40 digits)
100 μV
6 V
-6.000
to
6.000 V
± (0.05% of rdg + 3 digits)
± (0.1% of rdg + 15 digits)
1 mV
10 V
-10.000 to 10.000 V
Standard
signal
0.4-2 V
0.4000 to 2.0000 V
± (0.05% of rdg + 12 digits)
± (0.1% of rdg + 40 digits)
100 μV
1-5 V
0.800 to 5.200 V
± (0.05% of rdg + 3 digits)
± (0.1% of rdg + 15 digits)
1 mV
DI
Level
Threshold level (Vth = 2.4 V) accuracy ± 0.1 V
-
Contact
5
1 kΩ or less: 1 (ON), 100 kΩ or more: 0 (OFF)
(parallel capacitance of 0.01 μF or less)
-
rdg: Reading value
1
R, S, B, K, E, J, T, N: IEC60584-1, DIN EN60584, JIS1602
2
WRe3-25: W-3%Re/W-25%Re (Hoskins Mfg.Co.) ASTM E988
3
Pt100: JIS C 1604, IEC 60751, DIN EN60751
4
JPt100: JIS C1604, JIS C1606
5
DI contact detection current value: approx. 10 μA
6
Use standard mode to supply power through the USB connector.
Summary of Contents for GX70SM
Page 1: ...User s Manual IM 04L57B01 01EN 1st Edition Model GX70SM Wireless Input Unit User s Manual ...
Page 2: ......
Page 12: ...Blank ...
Page 148: ...Blank ...
Page 168: ...Blank ...