2-46
IM 04L57B01-01EN
13
Click
OK
.
An Information dialog box appears.
14
Click
OK
.
The calibration values will be saved to the GX70SM, and
disappears. While the values are
being saved, the GX70SM LED (green and red) blinks rapidly.
15
Click
Close
.
Note
If a calibration error occurs, “Exists” is shown next to Calibration error.
Operation complete
2.8 Calibrating the Universal Inputs and Built-in Humidity Sensor
Summary of Contents for GX70SM
Page 1: ...User s Manual IM 04L57B01 01EN 1st Edition Model GX70SM Wireless Input Unit User s Manual ...
Page 2: ......
Page 12: ...Blank ...
Page 148: ...Blank ...
Page 168: ...Blank ...