2.2 INSTALLATION OF ISC40 SENSOR IN ISC40FF
Installation of the flow fittings involves two steps:
•
Installation of the sensor
The sensor has to be mounted in the flow fitting. It is important that the position of
the sensor in the fitting allows easy flow through the hole of the sensor donut (refer
to Section 3 of ISC40 manual). Generally good flow is assured if the flats of the sensor
are oriented perpendicular to the outlet piping. It is also important that the sample
piping is oriented in such a way that the direction of the flow is upwards to assure
complete filling of the flow chamber
•
Installation of the fitting on wall, railing or stanchion
For this purpose, the flow fitting has an optional pipe/wall mounting kit /MS or /MP
(see Figure 1 and Figure 2 for details). This kit consists of a clamp ring with bolts and
nuts which clamps around the flow chamber. Therefore, the flow chamber can be
turned in the mounting assembly allowing more flexibility in installation. The mount
-
ing plate can be mounted on a wall or panel with 2-4 anchor bolts with a diameter
of 10 mm (3/8”) max. For those installations where pipe mounting is requested (2”
nominal pipe), either horizontal or vertical, a saddle and U-bolt have to be used.
Figure 1:
Pipe/wall mounting kit /MS
Figure 2:
Pipe/wall mounting kit /MP