4.  PART NAMES OF FRONT PANEL

The figure below shows the MA7D with its front panel (cover) 

being open.

 

For Output-1

For Output-2

Zero-adjustment volume
Span-adjustment volume

Zero-adjustment volume
Span-adjustment volume

Power indicator lamp

Turns on at power on.

5.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance. 

5.1  Calibration Apparatus 

 •

A DC voltage/current standard (Yokogawa GS200 or the 

equivalent)

 •

A digital multimater (Yokogawa DM7560 or the equivalent) 

 •

A precision resistor of 250 Ω ± 0.01%, 1 W 

5.2  Calibration Procedure 

(1) Connect the instruments as shown below. First adjust the 

output-1 signal and then the output-2 signal. 

(2) Use the DC voltage/current standard and apply input signals 

equivalent to 0, 25, 50, 75, and 100% of the input span to the 

product. Check to see the corresponding output voltages are 

0, 25, 50, 75, and 100% respectively and within the specified 

accuracy rating. “R” is used for current output. 

 • If the output signals are out of the accuracy rating range, 

adjust the output signal level using the zero and span 

adjustment volumes on front face of the product.

    

10 11

1 2

3

9

8 7

6 5

4

9

7

8

1

2

R

10

11

R

4

5

5

6

R: 250 

Ω 

precision resistor

     for current output

R: 250 

Ω 

precision resistor

     for current output

Power supply

Output-2

Input

With SINK current

Without SINK current

Output-1

2

1

2

1

2

1

L1

N2

GND

DC voltge/

current

standard

DC voltge/

current

standard

Digital

multimater

2

1

Digital

multimater

6.  TRANSPORT AND STORAGE CONDITIONS

 •

Temperature: -25 to 70°C

 •

Temperature change rate: 20°C per hour or less

 •

Humidity: 5 to 95%RH (no condensation)

 CAUTION

Keep this product in a conductive bag when plugged 

out, during transport or storage.

3

IM 77J04A07-02E