4.  PART NAMES OF FRONT PANEL

The figure below shows the MPT7 with its front panel (cover) 

being open. 

Zero-adjustment volume
Span-adjustment volume

Power indicator lamp
Turns on at power on.

5.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance. 

5.1  Calibration Apparatus 

 •

A AC voltage current standard (YOKOGAWA 2558A or the 

equivalent)

 •

A digital multimater (YOKOGAWA DM7560 or the equivalent)

 •

A precision resistor of 250 Ω ± 0.01%, 1 W

5.2  Calibration Procedure 

(1) Connect the instruments as shown below.

(2) Use the AC voltage current standard and apply input signals 

equivalent to 0, 25, 50, 75, and 100% of the input span to the 

product. Check to see the corresponding output voltages are 

0, 25, 50, 75, and 100% respectively and within the specified 

accuracy rating. “R” is used for current output. 

 • If the output signals are out of the accuracy rating range, 

adjust the output signal level using the zero and span 

adjustment volumes on front face of the product.

6 5

4 3

7 8

1 2

L+

N–

GND

6

7

8

1

2

R

+

3

4 ±

V

Power supply

Input

AC voltge 

current

standard

R: 250 

 precision resistor

for current output

DMM

Output

6.  TRANSPORT AND STORAGE CONDITIONS

 •

Temperature: −25 to 70°C

 •

Temperature change rate: 20°C per hour or less

 •

Humidity: 5 to 95%RH (no condensation)

 CAUTION

Keep this product in a conductive bag when plugged 

out, during transport or storage.