<4. WIRING>
4-4
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
WARNNING
The power sources for solenoid valves are provided internally. Use solenoid valves that have the
same power supply voltage and frequency specifications as the instrument.
4.4.4
Contact Output Wiring
All outputs are voltage-free, dry contacts (mechanical relay contacts) and rated 3A at 250 V AC or
30 V DC.
10
11
12
13
14
15
16
17
18
R3/HI
MAIT
・
CAL
FAIL
R2/LO COM
RANGE-OUT/HL ALM
CONTACT OUTPUT
F7.4E.ai
NO
CONNECTION
Higher-Level Devices, Annunciators, etc.
High/low limit alarm
contact output or
range contact output
Maintenance mode
contact output
FAIL contact output
Figure 4.4
Contact Output Wiring
(a) Wiring for High/Low Alarm Contact Output/Range Answerback Contact Output
Either high/low alarm contact output or range answerback contact output can be selected.
For their detailed functions and setting procedures, refer to Chapter 8, Parameter Settings.
Terminal 12 is a common terminal and the contact between terminals 10 and 12 and the one
between terminals 11 and 12 will open or close in response to the instrument status. For external
connections to higher-level devices or other instruments, refer to Figure 4.4. Both contacts are
open when the instrument is not powered.
Terminal No.
High/Low Alarm Contact Output
Range Answerback Contact Output
High Alarm
Low Alarm
Range 1
Range 2
Range 3
10-12
Closed
Open
Open
Closed
11-12
Closed
Open
Closed
Open
(b) Wiring for Maintenance Status Contact Output
The contact between terminals 13 and 14 is closed during maintenance. For its functions, refer to
Chapter 8, Parameter Settings. The contact is open when the instrument is not powered.