<6. OPERATIONS>
6-6
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
6.3.7
Functions Numbers
The functions assigned to the Function Numbers are listed in Table 6.3. In some functions
parameters are not shown unless the password is effective in MAINT mode.
Table 6.3
Function Numbers
Func
No.
Function
Setting and Range
Password
Not
Effective Effective
00
Entering password
01
Executing calibrations
ZERO key: executes zero calibration
SPAN key: executes span calibration
----
02
Setting calibration gas
concentrations
ZERO key: for zero gas concentration, 0 to 6
SPAN key: for span gas concentration, 0.5 to 25
03
Checking calibration
coefficients
ZERO key: displays zero calibration coefficient
SPAN key: displays span calibration coefficient
04
Setting output hold function
during calibration
0: enabled
1: disabled
05
Initializing calibration
coefficient
10
Setting remote range
switching contact input
1: disabled
2: enabled
11
Selecting range
1: range 1
2: range 2
3: range 3
12
Setting span for range 1
5 to 25% O
2
13
Setting span for range 2
5 to 25% O
2
14
Setting span for range 3
5 to 25% O
2
16
Fixed analog output
ZERO key: 4 mA
SPAN key: 20 mA
----
20
Setting low limit alarm
0 to 25% O
2
21
Setting high limit alarm
0 to 25% O
2
22
Setting high/low limit alarm
function
0: disabled
1: enabled
30
Checking temperature (°C)
of constant temperature
chamber
31
Checking sensor emf (mV)
40
Executing error check
----
41
Resetting sensor emf error
status
----
42
Resetting calibration
coefficient error status
----
43
Setting output hold function
when error occurs
1: enabled – output is held at last measured value
2: disabled
3: enabled – output is held at preset value
44
Setting preset value
–10 to 110%
45
Setting contact output
status when error occurs
0: contact closes when error occurs
1: contact opens when error occurs
50
Setting initial wait time for
automatic calibration
0 day 0 hour to 99 days 24 hours
51
Setting interval for
automatic calibration
0 day 1 hour to 99 days 24 hours
52
Setting zero calibration time 1 to 99 minutes
53
Setting span calibration
time
1 to 99 minutes
54
Setting purge time
1 to 99 minutes
55
Executing span calibration
0: not execute span calibration
1: executes span calibration
56
Setting the number of
times of skipping span
calibrations
0 to 99 times
57
Starting automatic
calibration function
0:
OFF
1: ON
58
Manual operations of
solenoid valves
ZERO key: opens solenoid valve in zero gas line
SPAN key: opens solenoid valve in span gas line
----
67
Checking error status
For service
98
Checking firmware version
99
Initialization
----
:
Data can only be viewed.
:
Data can be changed or functions can be executed.