< 2. Installation, Piping and Wiring >
2-7
IM 12B07N01-01E
2.4.2
Piping Procedure
(1) Air piping examples
Supply Air
Regulator
Drive Signal
Solenoid Valve*
1
To Wash Cycle Timer
Pressure 200 kPa (2 kg/cm
2
)
Minimum
Nominal 16 mm Wire-Reinforced Vinyl Tube*
2
Nominal 15 mm
Hard Pipe
Notes:
*1 If corrosive gas is produced at the measuring site, consider
using a corrosion-resistant solenoid valve.
*2 If the solution temperature is high, use a high temperature
resistant solenoid valve.
F2.11E.ai
Figure 2.11
Typical Brush/jet Cleaner Air Piping
(2) Industrial water piping examples
F2.12E.ai
Drive Signal
Solenoid Valve
Ball
Valve
To Wash Cycle Timer
Industrial Water
Pressure 200 kPa (2 kg/cm
2
)
Minimum
Nominal 16 mm Wire- Reinforced Vinyl Tube
Nominal 15 mm Hard Pipe
Figure 2.12
Industrial Water Piping for Typical Brush/Jet Cleaner
(3) Tap water piping examples
Drive Signal
Check
Valve
Nominal 16 mm
Wire-Reinforced
Vinyl Tube
*
F2.13E.ai
Pump
Drain
Float Valve
Tap Water
To Wash Cycle Timer
Cleaning
Water Tank
Note:
*: When using a Model PH8PU1 cleaner pump and tank, set the solution
pressure to 10 kPa (0.1 kg/cm
2
) or below.
Figure 2.13
Tap Water Piping for Typical Brush/Jet Cleaner