3-9
TI04L55B01-01EN
Event Data Recording Modes
The following event data recording modes are available.
l
Free
Recording starts when you start it and stops when you stop it.
Measurement data is divided by the specified duration (data length).
Time
File
File
File
Adding data
Data length
Set the size of recording data per file. Recording data is divided at the specified file size. The
selectable data lengths vary depending on the number of channels to record and the Scan
interval setting.
l
Single
The GM enters the trigger-wait state when you start recording. After a trigger event occurs, the
GM will record data for the specified time (data length) and stop. From this point, the GM will not
record even if the trigger condition is met.
Time
File
Trigger condition met
l
Repeat
The GM enters the trigger-wait state when you start recording. After a trigger event occurs, the
GM will record data for the specified time (data length) and stop. Then, it enters the trigger-wait
state again and repeats recording for the specified time (data length) every time the trigger
condition is met. To stop recording event data, stop the recording.
Time
File
File
Trigger condition met
Trigger condition met
3
Functions
Summary of Contents for SMARTDAC+ GM
Page 2: ...Blank Page...