3-29
TI04L55B01-01EN
User Privileges
Limitations can be applied to the operations that a user can perform.
Up to 10 conditions of operation limitations can be set (authority numbers 1 to 10).
For each user, you can set whether to apply user limitations and the conditions to be applied.
Operation limitations
Operation
Explanation
Recording
Restricts record start/stop operation.
Computation
Restricts math start, math stop, and math reset operations.
Data save
Restricts the operations to save display data, save event data, generate event
triggers, manual sample, snapshot, reset timer, and reset match time timer.
Message
Restricts message writing operations.
Batch
Restricts batch number writing, lot number writing, comment writing, and text field
writing.
Alarm ACK
Restricts alarm ACK operations.
Communication
Restricts the operations to start, stop, and test mail; test FTP, get and release
network information; test printer output; manually recover Modbus master; and
manually recover Modbus client.
Time set
Restricts manual SNTP time adjustment and time setting operations.
Setting operation
Restricts setting operations.
External storage medium Restricts the operations to save, load, and delete files; manually save data; save
alarms; abort saving; create certificate requests; install certificates; and install
intermediate certificates.
System operation
Restricts the operations to initialize; reconfigure system; manage, display, and
delete certificates; and execute unverified certificates.
Output operation
Restricts internal switch operations and relay operations.
Key Lock
The GM (Data Acquisition Module) keys (START, STOP, USER1, USER2) can be locked.
Lock
Key
GM10
3
Functions
Summary of Contents for SMARTDAC+ GM