14-4
IM 703155-01E
Operation Errors
Code
Message
Description and Corrective Action
Reference
Section
1000
Data is out of range.
Set it again.
–
1001
Illegal parameter value.
Set it again.
–
1002
Illegal character.
Set it again.
Chapters
10, 11
1003
Invalid argument.
Set it again.
–
1004
Cannot implement this function.
–
–
1005
Hardware missing.
Check the installed options.
–
1006
Password is wrong.
Enter the correct password.
–
1007
Invalid specified printer.
–
11.3
1008
Not found server.
Check the settings.
–
1020
Cannot be operated.
Attempted to carry out a prohibited operation.
–
1021
Cannot be operated while initializing.
Wait until the initialization is finished.
12.1
1022
Cannot be operated while calibrating.
Wait until it is finished.
12.1
1023
Cannot be operated while testing.
Cancel the test first.
14.3
1024
Cannot be operated while changing settings.
Wait until it is finished.
–
1025
Cannot be operated while accessing to media.
Wait until it is finished.
Chapter 10
1026
Cannot be operated while transmitting waveform
Wait until it is finished.
7.2
1027
Cannot be operated while analyzing.
Wait until it is finished.
–
1028
Cannot be operated while generating waveform.
Stop the waveform generation first.
6.2
1029
Cannot be operated while updating.
Wait until the update operation is finished.
–
1030
Cannot operate. Confirm Setting.
The setting is incorrect or the computation was in
–
error in the past.
1040
Cannot load this file that has invalid parameter.
Load the file again after eliminating the invalid
–
parameter.
1041
Number of waveform data points is less than 256
Set the waveform data with at least 256 points.
7.2,
points.
chapter 8
1042
Number of waveform data points exceeds the
Set the waveform data that does not exceed the
7.2
upper bound.
upper limit of data points.
1043
Cannot be registered to sequence list any further.
–
7.2
1044
All sequence list is empty.
–
7.1, 7.2
1045
Waveform data still not transmitted.
The waveform data has not been transmitted at all
7.2
to the waveform memory, or the sequence list has
been changed after transmission.
1046
Transmission size is too big at Output$.
Register the elements again so that the total number 7.2
of points of the elements is within the size that can
be transmitted to the waveform memory.
1047
Transmission error at No.### in Output$.
The ###th element in Output$ could not be transmitted
–
to the arbitrary waveform generation board.
Check the waveform data of the element that
caused the error.
1048
Transmission error at No.### in Output$.
Cannot open waveform data file.
7.2
Check whether the waveform data file of the
element that resulted in error is at the specified
location, or whether the file is corrupt.
1049
Transmission error at No.### in Output$.
Waveform data points is different from the length
7.2
registered in sequence list. Register the waveform
data file of the element that resulted in error again
to the sequence list.
1059
Waveform data file is not loaded.
Load the waveform data file.
10.3
1060
Waveform data size is too big.
Waveform data that exceeds the display analysis
–
memory size cannot be loaded.
14.2 Messages and Corrective Actions