2-10
IM 703155-01E
Start Trigger «For procedures, see section 6.1»
You can set the method of starting waveform generation.
Start Waveform Generation Using the START/STOP Key
Waveform generation starts when you press the START/STOP key on the front panel.
Press the START/STOP key again while waveform is being generated to stop waveform
generation.
Start Waveform Generation on a Start Trigger
When you press the START/STOP key on the front panel, the VB8300 enters the trigger-
wait state. Waveform generation starts when an external start trigger signal is detected
at the CLOCK START IN terminal on the rear panel. Press the START/STOP key while
waveform is being generated to stop waveform generation.
Waveform output
(analog signal)
CLOCK OUT
When the low-pass filter is set to Through
Timing chart when the external start trigger is used
EVENT OUT
CLOCK START IN
T1
T2
T3
T4
T1: 100 ns or more
T2: (1.5 to 2.5 clocks)
×
(1/clock frequency) ns + approx. 40 ns
T3: Approx. 22.5 ns
T4: Approx. 30 ns
Detection edge