3-3
IM 703155-01E
Instrument Preparation and Common Operations
3
3.2
Installing the Instrument
Installation Position
WARNING
To avoid the possibility of fire, never use the instrument with the rear panel
facing down. There are vent holes for the cooling fan on the rear panel. Placing
the instrument with the rear panel down can cause a fire when the instrument
malfunctions. If you must use the instrument with the rear panel down, place a
metal plate or a flame-resistive barrier (grade UL94V-1 or higher) beneath the
instrument.
• As shown in the figure below, place the instrument in a horizontal position or by using
the stand.
• When using the stand, pull the stand forward until it locks (perpendicular to the bottom
surface of the instrument). If you are not using the stand, return it to the original
position while pressing the leg section of the stand inward.
• If you are installing the instrument on a slippery surface, attach the rubber feet (four
pieces, included in the package) to the feet on bottom panel.
Place the instrument in a horizontal position or inclined position
using the stand.
Installation Conditions
Install the instrument in a place that meets the following conditions.
Well-Ventilated Location
Inlet/Vent holes are located on the top and bottom of the instrument. In addition, there
are vent holes for the cooling fan on the rear panel. To prevent internal overheating,
allow for enough space around the instrument (see the figure below) and do not block
the inlet and vent holes.
10 cm or more
VB8300
BASEBAND SIGNAL GENERATOR
5 cm or more
5 cm or more
10 cm or more