6-2
IM 703155-01E
6.2
Executing Waveform Output
Procedure
APPLI
START/STOP
ON/OFF
OUTPUT
Starting Waveform Generation
• Starting Waveform Generation Using the START/STOP Key
1.
Set the start trigger OFF according to the procedure given in section 6.1.
2.
Press
START/STOP
. The green indicator above the START/STOP key
illuminates, and waveform generation starts.
• Starting Waveform Generation on a Start Trigger
1.
Set the start trigger ON according to the procedure given in section 6.1.
2.
Press
START/STOP
. The green indicator above the START/STOP key illuminates.
When conditions for accepting the trigger* are met and an external start trigger
signal is detected, waveform generation starts.
*
For details on the trigger acceptance conditions, see section 2.5.
Note
If you wish to output the waveform data that is loaded in the waveform memory from the beginning,
press the ON/OFF key to turn the waveform output ON* and start waveform generation.
*
The green indicator above the ON/OFF key illuminates.
Stopping Waveform Generation
• Manually Stopping Waveform Generation
3.
Press
START/STOP
while waveform generation is in progress (while the green
indicator above the START/STOP key is ON). Waveform generation stops.
• Stopping after the Waveform Data Loaded in the Waveform Memory Is Output
Waveform generation stops when the waveform data loaded in the waveform memory
is output to the end. However, if the you set the repeat count to infinity when you
edited the sequence, waveform generation continues until you manually stop it.
Explanation
Starting/Stopping Waveform Generation
Waveform generation can be started simultaneously on all output terminal sets (OUT1 to
4). The green indicator above the START/STOP key illuminates while waveform
generation is in progress.
Turning the Waveform Output ON/OFF
Controls the signal output ON/OFF. If the waveform output is not turned ON, waveform
will not be output even if waveform generation is started.
• When the waveform output is turned ON, the green indicator above the ON/OFF key
illuminates. If you press the ON/OFF key while the indicator is illuminated, waveform
output is turned OFF.
• When the waveform output is OFF, the output terminal is set to ground level through
the 50-
Ω
impedance.
• For the timing relationship between waveform generation and waveform output, see
section 2.5.
• The digital output in independent of the waveform output ON/OFF setting.