8-20
IM 703155-01E
• Horizontal Display Start and End Points (X Begin Value/X End Value)
You can set the minimum value (the scale value at the left edge of the waveform
display frame) and the maximum value (the scale value at the right edge of the
waveform display frame) of the amplitude of the instantaneous power to be displayed.
Range: 0.00 to 50.00 (in 0.01 steps)
Be sure that (display end point – display start point) is greater than 0.
Default values: X Begin Value 0.00, X End Value: 5.00
• Horizontal Axis Auto Scale (X Auto Scale)
The horizontal scale is set so that the amplitude of the instantaneous power that is
calculated from the waveform data fits on the waveform display frame.
Y Top Value
X Begin Value
X End Value
1.0E 0
1.0E–1
1.0E–2
1.0E–3
1.0E–4
Display example when
Y Top Value = 1
Y/Div = 4
Divided into
the specified
number of
orders of
magnitude
specified by
Y/Div.
Marker Measurement
Marker Type
You can select the waveform from below.
OFF
The marker is not shown on the display, and you cannot set the marker.
Marker
The amplitude of the instantaneous power at the marker position and the accumulated power
of the instantaneous power that exceeds the average power can be displayed.
Marker
Output terminal set
Waveform data file name
The amplitude of the
instantaneous power at
the marker position and
the accumulated power
of the instantaneous
power that exceeds the
average power
8.3 CCDF Display