15-6
IM 703155-01E
15.9 General Specifications
Item
Specifications
Standard operating conditions
Ambient temperature:
23
±
2
°
C
Ambient humidity:
55
±
10% RH
Error in supply voltage and frequency: Within 1% of rating
Warm-up time
At least 30 minutes
Storage environment
Temperature:
–20 to 60
°
C
Humidity:
20 to 80%RH (no condensation)
Operating conditions
Temperature:
5 to 40
°
C
Humidity:
20 to 80%RH (no condensation)
Storage altitude
3000 m or less
Operating altitude
2000 m or less
Rated supply voltage
100 to 120 VAC/220 to 240 VAC (automatic switching between 100-V/200-V systems)
Permitted supply voltage range
90 to 132 VAC/198 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Permitted supply voltage frequency range
48 to 63 Hz
Power fuse
250 V/5 A time lag, VDE/SEMKO/UL/CSA/SEV certified
Maximum power consumption
300 VA
Withstanding voltage
1.5 kVAC for one minute.
(between power supply and case)
Insulation resistance
500 VDC, 10 M
Ω
or more
(between power supply and case)
External dimensions
426 mm (W)
×
176 mm (H) (W)
×
450 mm (D),excludes the handle and other
(details on the next page)
projections
Weight
Approx. 16 kg (for Model 703155-128-8/D8/AT)
Instrument’s cooling method
Forced air cooling
Installation position
Horizontal (however, the stand can be used)
Battery backup
Setup data and clock are backed up with the internal lithium battery
Battery life:
Approx. 3 years (at ambient temperature of 23
°
C with the power
turned OFF)
Recommended calibration period
1 year
Safety standards
Complying standard
EN61010-1
• Overvoltage category II
1
• Pollution degree 2
2
Emission
Complying standard
EN61326 Class A, C-Tick AS/NZS CISPR 11
This product is a Class A (for commercial environment) product. Operation
of this product in a residential area may cause radio interference in which
case the user is required to correct the interference.
Cable condition
• BNC input/output terminal
Use a BNC cable of length 2 m or less.
• Video output connector
Use a 15-pin D-Sub VGA shielded cable of length 2 m or less.
• RS-232 connector
Use a RS-232 shielded cable of length 3 m or less.
• IO digital output connector
Use the IQ digital output cable (A1388WL) that comes with the package.
• Ethernet connector
Use a Category 5 or better shielded cable of length 30 m or less.
• USB connector
Use a USB shielded cable of length 3 m or less.
Immunity
Complying standard
EN61326 commercial environment
Influence in the immunity environment
No influence to the output waveform
Cable condition
Same as the cable condition for emission
1 Overvoltage Categories define transient overvoltage levels, including impulse withstand voltage levels. II applies to electrical
equipment that is powered by a fixed installation such as a distribution board.
2 Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution Degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).