<4. OPERATION>
4-7
IM 12E04A02-02E
<Time chart>
SV1
(Draining)
SV3
(Water sample)
SV4
(Zero-turbidity
reference water)
OFF
OFF
OFF
OFF
ON
ON
OFF
ON
N =
1
2
3
4
ON
HOLD ON
Cleaning start
Cleaning time 2
One cleaning cycle
Recovery time
Drain time 2
Water fill time 2
Water fill time 2
N (number of cleaning repetitions) = 4 (example)
SV2
(Cleaning)
Output
holding *1
F0408.ai
*1 During cleaning, the output is always in the hold state.
The display is also set to be held using FUNCTION "8." in the <PROGRAM1> mode.
Contact outputs C1 and C2 can be used as contacts for the period during cleaning and
calibration. This is set using FUNCTION "E." in the <PROGRAM1> mode.
Note 1:
N: Number of cleaning repetitions
Note 2:
One automatic cleaning cycle = N (dr f cleaning) + (dr f recovery)
Figure 4.8 Automatic-cleaning Operation Time Cheat (Suf
fi
x code: -A3)
Table 4.3 Setpoints for Automatic Cleaning and Automatic Zero Calibration
Item
Setting
range
Factory
setting
Unit
Setting mode
Mode
FUNCTION
Cleaning period
0.1 to 24.0
2.0
Hours
PROGRAM 2
1
Cleaning calibration ratio
1 to 10
1
times
PROGRAM 2
2
Cleaning time 2
10 to 120
30
Sec
PROGRAM 2
5
Draining time 2
1 to 20
10
Sec
PROGRAM 2
7
Water
fi
lling time 2
10 to 120
100
Sec
PROGRAM 2
9
Relaxation time
10 to 120
150
Sec
PROGRAM 2
A
Number of cleaning repetitions
30 to 600
5
times
PROGRAM 2
3
For the setting procedures for each setpoint, see Section 6.7.
3rd Edition : May. 31, 2010-00
Summary of Contents for Vigilant Plant EXA TB Series
Page 49: ...Blank Page ...
Page 59: ...Blank Page ...
Page 119: ...Blank Page ...
Page 125: ...Blank Page ...
Page 127: ...Blank Page ...
Page 133: ...Blank Page ...