<6. PROCEDURES FOR CONVERTER OPERATION>
6-5
IM 12E04A02-02E
Table 6.5 Functions in <PROGRAM1> Mode
F
Operation
Setting range/
selection
Initial
indication
Remarks
1 Output range 1 setting
0.00 to 2000
mg/l (deg.)
L00.00
H02.00
2 Output range 2 setting
0.00 to 2000
mg/l (deg.)
L00.00
H020.0
3 Output range 3 setting
0.00 to 2000
mg/l (deg.)
L00.00
H0200
4 Average coef
fi
cient setting (time constant
setting)
1 to 255
00030
The 30 to 90 % response
time is about 22 seconds.
5 Setting of output value in a failed condition
-10 to 110 %
00000
6 Output setting of line-segment approximation
0 to 100 %
00050
7 Auto-range changeover point setting
70 to 100 %
00080
8 High alarm setting
-010.0 to 2200.0
mg/l (deg.)
2200.0
9 Low alarm setting
-010.0 to 2200.0
mg/l (deg.)
-010.0
A Setting of turbidity value by calibration plate
0000.0 to 2000.0
mg/l (deg.)
0090.0
This varies with the
calibration plates.
B Setting of zero-correction factor
-9.000 to 09.000
mg/l (deg.)
00.000
C Setting of sensitivity correction factor
0.2500 to 4.0000
1.0000
D
E
F
1. Auto-range selection *1
ON/OFF
OFF
ON: Execution
OFF: Stop
2. Remote range selection *1
ON/OFF
OFF
ON: Execution
OFF: Stop
3. Local range selection *1
_ _ 1/_ _ 2/_ _ 3
_ _ 1
1: Output range 1
2: Output range 2
3: Output range 3
4. Output hold setting in the <MAINT.>,
<PROGRAM 1>, or <PROGRAM 2> mode
ON/OFF
ON
ON: Hold
OFF: No-hold
5. Output mode setting in failure occurrence
_ _ 1/_ _ 2/_ _ 3
_ _ 1
1: No-hold
2: Hold
3: Preset
6. Setting to output hold for 4 mA or less
ON/OFF
OFF
ON: Execution
OFF: Stop
7. Setting to hold negative turbidity display
ON/OFF
OFF
ON: Execution
OFF: Stop
8. Setting to display hold in automatic cleaning
or automatic calibration
ON/OFF
OFF
ON: Execution
OFF: Stop
9. Basic code (sampling speci
fi
cations) setting
_ _ 1/_ _ 2/_ _ 3
_ _ 3
1: -NN, -A1
2: -A2
3: -A3
A. Automatic cleaning function execution/stop
ON/OFF
ON
ON: Execution
OFF: Stop
B. Automatic calibration function execution/stop
ON/OFF
ON
ON: Execution
OFF: Stop
C. Selection of unit to indicate
_ _ 1/_ _ 2/_ _ 3
_ _ 1
1: mg/l
2: Degree
3: FTU *2
D. Selection of calibration plate/standard
solution
ON/OFF
ON
ON: Calibration
plate
OFF: Standard solution
E. Selection of AUX cotact output
ON/OFF
ON
ON: Upper and lower limit
alarm
OFF: In automatic cleaning
or automatic calibration
F.
(*1)
The priority of these items are as follows (see subsection 5.2.1):
Auto-range > Remote range > Local range
(*2)
"FTU" unit, a turbidity unit when formazin standard is used, is also available. Since the instrument uses kaolin standard,
reference sensitivity calibration using formazin standard is required for the FTU display.
Note 1:
The values in the initial value column are those given when the memory storing the data is initialized. These may be partially
different from the values set upon shipment from the factory.
Note 2:
If an unused function code which is selected, "_ _ _ _ _" is displayed.
Note 3:
If a value out of the set range is entered, "not" is displayed. Pressing the [>] key restores this indication to the data display.
3rd Edition : May. 31, 2010-00