7-7
IM 04P02B01-01E
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
7
Explanation
RJC
Sets the RJC mode.
Internal: Uses the RJC function on the recorder.
External: Uses an external RJC function. If external is selected, set the compensation
voltage.
Volt
Sets the compensation voltage when using an external RJC function.
The compensation voltage can be set in the range of –20000
µ
V to 20000
µ
V.
7.4 Setting the RJC Function on Channels Set to TC Input