9-16
IM 04P02B01-01E
9.8
Setting the Partial Expanded Recording
This section explains the details of expanding a section of the recording range.
If you change the computing equation or the recording span, set the partial expanded
recording again.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key or
SHIFT
+
key to select
Math
and then press the
key.
3.
Press the
key or
SHIFT
+
key to select
Aux
and then press the
key.
4.
Press the
key or
SHIFT
+
key to select
Partial
and then press the
key.
5.
Set each item and press the
key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the
Setting complete
screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the
ESC
key.
7.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Set the channel range.
The new setting takes effect.
ESC/?
Select On to use the partial expanded
recording function.
Set the boundary position as
a percentage where the recording span
is assumed to be 100%.
Set the boundary value within the
recording span.
First channel
Last channel
0A-0A Channel
Setting complete
Set=Math
Math=Aux
Aux=Partial
CH=0A-0A
Partial=Off
Expand=50
Boundary= 0.00
<Related Topics>
Enabling the partial expanded recording function: Section 7.13