4. Installation
4.1 Precautions
Before installing XS550, read the cautionary notes in subsection 2.4 “Selecting
the Installation Location”. For additional information on the ambient conditions
allowed at the installation location, refer to chapter 8 “General Specifications”.
IMPORTANT
Connector Protection
To protect the connector, use a protective cap (Part Number: F9097ND)
when replacing batteries or installing without XS110A.
The protective cap is used to temporarily protect the XS550 connector
when the XS110A is removed from the XS550 during battery replacement
and other operations. When the XS110A is removed, waterproof and
dustproof performance cannot be guaranteed. Therefore, install the
XS110A on the XS550 as soon as possible after work.
Installation Work
• When welding piping during construction, take care not to allow welding
currents to flow through XS550.
• Do not step on this instrument after installation.
Mounting with XS110A
After installing the XS110A, be sure to check that the head of the XS110A
fixing screw of the XS550 has been pulled out to a level with the surface of
the XS110A and that the measuring module fixed screw of the XS110A is
properly tightened.
4.2 Mounting
This product can be attached to a 2-inch pipe using a mounting bracket. Refer
to Section 4.2.1 "mounting the XS110A" for mounting the XS110A. For details,
refer to Section 4.2.2 "Mounting using mounting brackets (When using a U
bolt)" or Section 4.2.3 "Mounting using mounting brackets (When using a belt)".