50
⚫
High-level disinfection
For semi-critical use of transviginal probes, high-level disinfection methods must be used.
a)
Immerse the probe in a 0.55% Phthalaldehyde disinfectant solution, and soak it at
room temperature (18
~
26
℃
) for 12
~
45 minutes. During the disinfection process, it
must be ensured that the liquid level does not exceed the fixed posts of the puncture
frame on both sides of the probe.
Intracavity probe immersion level
b)
After disinfection, immerse the probe in sterile water to clean and remove residual
disinfectant. It needs to be soaked in sterile water 3 times, 1 minute each time, and
replace with new sterile water after each soaking. During the cleaning process, make
sure that the socket part of the probe is out of the liquid.
c)
After cleaning, dry the probe with a sterile cloth or gauze, or place it in a well-
ventilated area to let it dry completely. Do not dry the probe.
d)
Check the probe carefully. Only the probe that is in perfect condition can be used.
The specific inspection method is described in accordance with the guidelines in
7.3.1.
e)
Precautions for disinfectant: For dilution and concentration, disinfection methods,
and use process, please refer to the instructions for disinfectant concentration and
disinfection method in the instructions provided by the manufacturer.
Declaration: the above cleaning and disinfection methods have been confirmed
by professional testing institutions, which is safe and effective. The operator can
rest assured to use.
WARNING
:
1.
It is strictly prohibited to immerse the handle and probe (except
the front end of the probe in the cavity) into any type of liquid or
detergent.
2.
No liquid of any kind is allowed to enter the handle and probe.
3.
It is strictly forbidden to disinfect the handle and probe with gas
or heating.
Fluid level