52
8 Puncture guide
8.1 Enter or exit puncture mod
⚫
Under the Windows platform, in the SoftKey area, click B mode [Image Parameter Settings],
and select [puncture (on / off)].
Set [Bracket] for quick calibration.
Set [guide line], depending on needle selection, A, B, and ALL.
Set [linetype], have thick, medium, thin.
⚫
Exit puncture guiding function
:
⚫
The system is in the puncture status. Click[Exit] to exit the puncture status.
8.2 Puncture lead calibration
Put the biopsy guide and transducer Acoustical lens under water.
Make sure the rubber of the transducer is above the water.
Click [calibrate] to align the needle with the puncture guide line by adjusting the “position” “angle”
“+” and“-”.
After coincident, save the calibrated puncture guide wire parameters.
WARNING
:
1. The guide leads must be calibrated before each puncture.
2. The frozen is not allowed at the time of puncture.
3. If the puncture needle is not consistent with the position of
the puncture guide lead, do not puncture.