2. Low limit setting--It will display the set weight low limit and the signal “-LL-” circularly when
you begin to set it. The operation is the same as setting high limit. It will go into the alarm method
setting after the low limit setting finished.
3. Alarm method setting—It means in the range alarm when the scale displays “-IN-”. It means out
of the range alarm when the scale displays “-OUT-”. It means no alarm when the scale displays
“-NO-”. You may press
to switch. Press
Enter
to confirm.
Press
Set to exit from the setting state when in the process of setting. The set value is invalid.
Counting model
Press Set to sample weight in the counting model. It displays “SAP X” (X is the sample number).
1.Press
to choose samples number from 10, 20, 50,100 and 200. Press
Enter
Enter
Enter
Enter
to make sure.
2.The weight window displays “LOAD-C” after the sample number is chosen. Put enough number
of samples then press
Enter
Enter
Enter
Enter
to finish sample.
3.There are two instance of unit weight lacking:
a.
When the weight window displays “-LAC-”, it means the unit weight of sample is less than
80% of division. You still can count but the counting may be inaccurate. The signal will
disappear after about 3 seconds.
b.
When the weight window displays “-CSL-”, it means the unit weight of sample is not
enough at all. Please press
Set
Set
Set
Set
to re-sample. Press
Enter
Enter
Enter
Enter
to exit back to the counting
model.
Press
Set
Set
Set
Set
to exit from the sample state in the process of sample.
Percentum model
1.Press
Set
in the percentum state to display “LOAD-P”. There are two ways to sample:
a.
Put enough samples on the platform then press
Enter
to finish sample.
b.
Press Unit then the scale displays “000000”. Press
and
Enter
Enter
Enter
Enter
to input the weight
value.
2.When the sample weighed less than 0.1% FS., the weight window displays “-CSL-”. It means
the sample is not enough. You should re-sample. Press
Set
Set
Set
Set
to re-sample. Press
Enter
Enter
Enter
Enter
to exit
back to the weighing model.
Press
Set
Set
Set
Set
to exit from the sample state in the process of sample.
Calibration
Keep pressing
Cali.
Cali.
Cali.
Cali.
for about 4 seconds. Calibrate after the weight window displays “CAL.”.
There are two ways to calibrate:
1.
Full range load: Put enough weight on the platform according to the full scale. Press
Enter
Enter
Enter
Enter
to make sure. Calibration finished.
2.
Random load:
a.
Press
Set
Set
Set
Set
, the weight window displays the weight calibrated last time.
b.
Press
and
Enter
Enter
Enter
Enter
to input the weight you want to load. Then load the corresponding
weight. Press
Enter
Enter
Enter
Enter
to make sure. Calibration finished.
Press Model
Model
Model
Model
to exit from the calibration state in the process of calibrating.
Note:
The high limit and low limit set in the weighing model are saved in the scale.
Different unit corresponding to different alarm range and method. The value inputted
by pressing
cannot be more than the full scale.
IV. Alarm indication:
1.
The weight window displays “--OF--” with continuously alarm sound when the weight
over 100%FS+9d. If the ADC overflows, the weight window displays “—Adc—”with
continuously alarm sound. For these instances, the weight should be taken away.
2.
If the voltage of the accumulator is low, the weight window displays “—Lo—” while the
weight is zero (The weight display will recover to normal when it is loaded ). Under this
condition, you can use the scale in a short time, but you should plug in the AC plug as