2. Low limit setting--It will display the set weight low limit and the signal “-LL-” circularly when 
you begin to set it. The operation is the same as setting high limit. It will go into the alarm method 
setting after the low limit setting finished. 
3. Alarm method setting—It means in the range alarm when the scale displays “-IN-”. It means out 
of  the  range  alarm  when  the  scale  displays  “-OUT-”.  It  means  no  alarm  when  the  scale  displays 
“-NO-”. You may press

to switch. Press

Enter

to confirm. 

  Press

Set to exit from the setting state when in the process of setting. The set value is invalid. 

 

Counting model

 

Press Set to sample weight in the counting model. It displays “SAP X” (X is the sample number). 
1.Press

to choose samples number from 10, 20, 50,100 and 200. Press

Enter

Enter

Enter

Enter

to make sure. 

2.The weight window displays “LOAD-C” after the sample number is chosen. Put enough number 
of samples then press

Enter

Enter

Enter

Enter

to finish sample. 

3.There are two instance of unit weight lacking: 

a.

 

When the weight window displays “-LAC-”, it means the unit weight of sample is less than 
80%  of  division.  You  still  can  count  but  the  counting  may  be  inaccurate.  The  signal  will 
disappear after about 3 seconds. 

b.

 

When  the  weight  window  displays  “-CSL-”,  it  means  the  unit  weight  of  sample  is  not 
enough at all. Please press

Set

Set

Set

Set

to re-sample. Press

Enter

Enter

Enter

Enter

to exit back to the counting 

model. 

Press

Set

Set

Set

Set

to exit from the sample state in the process of sample. 

 

Percentum model

 

1.Press

Set

in the percentum state to display “LOAD-P”. There are two ways to sample: 

a.

 

Put enough samples on the platform then press

Enter

to finish sample. 

b.

 

Press Unit   then the scale displays “000000”. Press

and

Enter

Enter

Enter

Enter

to input the weight 

value.

 

2.When  the  sample  weighed  less  than  0.1%  FS.,  the  weight  window  displays  “-CSL-”.  It  means 
the sample is not enough. You should re-sample. Press

Set

Set

Set

Set

to re-sample. Press

Enter

Enter

Enter

Enter

to exit 

back to the weighing model. 
    Press

Set

Set

Set

Set

to exit from the sample state in the process of sample.

 

 

Calibration 

 

Keep pressing

Cali.

Cali.

Cali.

Cali.

for about 4 seconds. Calibrate after the weight window displays “CAL.”. 

There are two ways to calibrate: 
1.

 

Full range load: Put enough weight on the platform according to the full scale. Press

Enter

Enter

Enter

Enter

to make sure. Calibration finished. 

2.

 

Random load:   

a.

 

Press

Set

Set

Set

Set

, the weight window displays the weight calibrated last time. 

b.

 

Press

and

Enter

Enter

Enter

Enter

to input the weight you want to load. Then load the corresponding 

weight. Press

Enter

Enter

Enter

Enter

to make sure. Calibration finished.

 

 Press Model

Model

Model

Model

to exit from the calibration state in the process of calibrating. 

Note: 
  The high limit and low limit set in the weighing model are saved in the scale. 
Different unit corresponding to different alarm range and method. The value inputted 
by pressing

cannot be more than the full scale. 

 

IV. Alarm indication: 

1.

 

The  weight  window  displays  “--OF--”  with  continuously  alarm  sound  when  the  weight 
over  100%FS+9d.  If  the  ADC  overflows,  the  weight  window  displays  “—Adc—”with 
continuously alarm sound. For these instances, the weight should be taken away. 

2.

 

If the voltage of the accumulator is low, the weight window displays “—Lo—” while the 
weight is zero (The weight display will recover to normal when it is loaded ). Under this 
condition,  you  can  use  the  scale  in  a  short  time,  but  you  should  plug  in  the  AC  plug  as 