Checking Your Blood Glucose
80
7
If control solution results are out of this range, REPEAT THE TEST.
Out-of-range control solution results
Out-of-range test results may be caused by:
•
Expired or bad control solution
•
Expired or bad test strip
•
Error in performing test
•
Code on test strip vial does not match code set in the PDM
•
Malfunction of the System
•
Control solution test done outside 59° to 104° F (15° to 40° C)
If your control solution test results continue to fall outside the
range printed on the test strip vial:
1. The System may not be working properly.
2. Do not use the System to test your blood glucose.
3. Call Customer Care.
■
Performing a Control Solution Test
1. Grasp the bottom of the test strip with the printed side facing
up (Figure 7-2).
2. Insert the top of the test strip into the test strip port (Figure 7-3
on the next page) until it stops. This turns on the PDM and dis-
plays the code number on the screen. Refer to page 54 to
check Pod status and to
Confirm ID
.
Figure 7-2
Top, to insert into meter
FreeStyle® test strip
Bottom edges, for blood sample