SECTION 3
MAKING MEASUREMENTS
The YSI Model 30 is factory calibrated. This means that once you have put batteries in the
instrument, you are ready to begin taking readings immediately.
The following diagram is an overview of the operation of the Model 30 and 30M. Note the
difference in key strokes between some of the 30 and 30M functions. See the following pages for
details of operation.
3.1
TURNING THE INSTRUMENT ON
Once the batteries are installed correctly, press the ON/OFF key. The instrument will activate all
segments of the display for a few seconds, which will be followed by a self test procedure which
will last for several more seconds. During this power on self test sequence, the instrument’s
microprocessor is verifying that the instrument is working properly and will display the cell
constant of the probe when the self test is complete. If the instrument were to detect a problem, a
continuous error message would be displayed. See Section 9, Troubleshooting for a list of error
messages.
After the self test procedure is complete, the temperature will be displayed in the lower right of the
display and the instrument is ready to make a measurement. Simply insert the probe into the liquid
sample about which you would like to receive information. It is important that the probe be inserted
into the liquid deep enough so that the hole on the side of the probe is completely covered by the
liquid. If possible, do not allow the probe to touch any solid object while you are taking readings;
Summary of Contents for 30
Page 2: ......
Page 4: ...ii ...
Page 6: ...Introduction Section 1 YSI Incorporated Model 30 30M 2 ...
Page 16: ...Saving Data Section 4 YSI Incorporated Model 30 30M 12 ...
Page 24: ...Advanced Setup Section 6 YSI Incorporated Model 30 30M 20 ...
Page 28: ...Principles of Operation Section 7 YSI Incorporated Model 30 30M 24 ...
Page 38: ...Warranty and Repair Section 10 YSI Incorporated Model 30 30M 34 ...
Page 40: ...Warranty and Repair Section 10 YSI Incorporated Model 30 30M 36 ...
Page 42: ...Accessories and Replacement Parts Appendix B YSI Incorporated Model 30 30M 38 ...
Page 48: ...Conversion Chart Appendix D YSI Incorporated Model 30 30M 44 ...