background image

Negative pressure regulating

Inspection & test on the overflow device

Note:  Dredge  the  suction  catheter  if  blocked  as  per  the  following  method: 
Bend  the  suction  conductor  in  "V"  form  (with  no  liquid  in  the  holder),  and 
release  it  to  the  original  status  when  the  negative  pressure  reaches  up  to 
the maximum value. Repeat this procedure several times till the catheter is 
not blocked.

Block the suction inlet, open the aspirator switch and regulate the negative 
pressure valve, and the readings on the pressure meter shall be within 0.02 
MPa ~ limit negative pressure.
Control  the  negative  pressure  as  required  for  suction  by  means  of  the 
negative pressure valve at the time of clinical practice.
Increase the negative pressure by turning the valve clockwise.
Reduce the negative pressure below 0.02 MPa prior to power shut-off.

Turn tightly the negative pressure regulating valve clockwise, and block the 
air suction inlet with the finger or the rubber head of dropper, or fold up and 
hold the suction tube.
Start  the  aspirator  for  running  with  no  strange  sound;  the  pointer  of  the 
vacuum meter will quickly reach up to the limit negative pressure. Release 
the  air  suction  inlet,  the  pointer  will  return  below  0.02  Mpa.  If  so,  the 
connector can be regarded as being in good connection.
Attach the phlegm suction catheter. The negative pressure in the negative 
pressure  system  shall  be  less  than  0.07  MPa  at  the  time  of  attaching  F8 
suction catheter, less than 0.03MPa when attaching F12 suction catheter. If 
so, the phlegm aspirator is considered as being in normal condition.

Connector inspection

4

Power

 

line

 

connection

Connect the plug with the power source. Turn on the power supply, and the 
power indicator will illuminate.
Note:  The  power  plug  is  used  for  power  shut-off,  and  the  power  socket 
shall be grounded reliably.       

Note: 
1. The liquid level still continuously ascends after the overflow device has 
been shut off, possibly due to:
(1) Residual negative pressure still in the holder.
(2) Valve mouth not fully closed.
For Item (1), the liquid level in the holder will not ascend when the suction 
tube conductor is placed again into the liquid as suctioned, and for Item (2), 
the liquid level still ascends. Thus, it is required to observe carefully, and lift 
immediately  the  conductor  out  of  the  suctioned  liquid  when  the  holder  is 
close to full, then, switch off the aspirator to stop suction, and examine the 
possible reason of the valve fault.
2.  The  float  is  still  adhered  on  the  valve  mouth  as  already  closed  by  the 
float,  possibly  due  to  the  negative  pressure  in  the  line.  At  this  moment, 
release  the  regulating  valve  or  shut  off  the  aspirator  (to  release  the 
negative pressure in the line), the float will descends from the valve mouth 
under  the  action  of  gravity.  (It  is  forbidden  to  pull  the  float  with  hand,  in 
order to avoid the rubber valve clack being separated from the float).
After  shut-off,  release  the  negative  pressure,  then,  open  the  holder  plug;
Never  use  the  aspirator  under  the  condition  of  the  overflow  device  &  the 
conductor dismantled.

Open  the  holder  plug;  clean  up  the  valve  mouth,  and  leveling  the  rubber 
valve  clack  on  the  float.  The  valve  clack  shall  not  be  warped,  bent  and 
broken,  but  well  connected  with  the  float.  The  float  shall  be  able  to  move 
freely in its support without any blockage, lift the holder plug with hand to 
make  the  float  contact  the  water  surface  perpendicularly  gradually  lower 
the holder cover to let the float rise.
Tighten  the  hold  plug,  attach  the  suction  tube  conductor  at  the  inlet,  and 
screw firmly the regulating valve, then, actuate the aspirator.
Put the suction conductor into one clean water pail or attempt to simulate 
actual application to suction the liquid into the holder of the overflow device. 
As a result, the float will rise as the liquid level ascends until the valve is closed 
and suction stops automatically. The final position of liquid level depends on the 
suction process adopted.
Release the regulating valve, set the aspirator switch off, open the holder plug 
and empty the liquid in the holder. The float shall be at the bottom of the support 
and the valve is in open status in case of re-screwing firmly the hold plug.
If so, the overflow device is considered as being in normal condition, which can 
be used for clinical practice.

5

Summary of Contents for 7E-B5

Page 1: ...Phlegm Suction Unit JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Sales Center Address Huanyuan East Road No 1 Xuzhuang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address Yu...

Page 2: ...only describes the usage of the product Do not use accessories other than the manufacturer s recommendation This will reduce the performance of the machine 4 Please do not use the machine and return...

Page 3: ...agram 2 Product Features Installing and Commissioning The customer shall carefully inspect if the appearance of product is good and the varieties quantities of the attachments are in conformity with t...

Page 4: ...he holder will not ascend when the suction tube conductor is placed again into the liquid as suctioned and for Item 2 the liquid level still ascends Thus it is required to observe carefully and lift i...

Page 5: ...in the air filter which leads to gradually darkening of the color of filter diaphragm and obviously reducing or even disappearing of suction force at the inlet of tube while the negative pressure indi...

Page 6: ...ng and application Wipe the case outer surface with lightly wet rag already soaked in the disinfectant and prevent any liquid seeping into the pump Never wipe the places marked with letters and patter...

Page 7: ...Operator due to electromagnetic disturbances Attachments 11 10 1 Suction conductor length 2m 7 12 1 pc 2 Suction catheter F8 F12 1 pc respectively of child adult 3 Power cord 1 pc 4 User s manual 1 pc...

Page 8: ...customer or the user should assure that it is used in such environment Guidance and manufacturer s declaration electromagnetic immunity 3 A m Power frequency magnetic fields should be at levels charac...

Page 9: ...etic environment in which radiated RF disturbances are controlled The customers or the users of these YUWELL aspirators can help prevent electromagnetic interference by maintaining a minimum distance...

Reviews: