Limit negative
pressure<
0.08MPa
1)Holder mouth leakage
2)Leakage on
connecting points
3)Regulating valve
loose or released
4)Surrounding
atmosphere is not as
required.
1)Remove dirt, tighten or
change the holder cover,
seal ring, and connector
2)Re-tighten each connection
point
3)Turn tightly the regulating
valve
4)Move the machine to the
required atmosphere
Change the
broken
suction
catheter
Solution
Remark
Probable reasons
Problem
Trouble shooting
9
8
Maintenance
It is recommended to have the suction tube suctioned small amount of
clean water for cleaning up the inner wall before switching off the aspirator.
After use, empty the holder, clean up dirt on the holder and plug with soft
brush or rag, flush it with water and conduct sterilization. (including the
overflow device, the seal ring and various tubes. Unscrew the overflow
device, and separate the float from its support for completely cleaning up, if
necessary. (Note: The rubber valve clack shall not be separated from the
float.)
Use the physiological saline to clean out the residual strong phlegm and
mucus in the tube after used. Replace the suction catheter if not smooth. It
is recommended to adopt one-time suction catheter.
Place the holder, cover and all tubes into the disinfectant compounded with
the KONVIDA disinfector tablets (0.5 g per tablet) in 1:500 concentration for
1 hour
Note: Keep the holder away from any sharp utensils to avoid drop in the
process of cleaning and application.
Wipe the case outer surface with lightly wet rag already soaked in the
disinfectant, and prevent any liquid seeping into the pump. Never wipe the
places marked with letters and patterns.
Place
the machine in dry and clean places, and periodically start running once a
time
(normally one time every 6 months).
Note
: Install the overflow device, conductor and other tubes as per the
connecting
mode before re-use.
Note: The dismantling & repair on the pump body if fault shall be conducted
by the specialized worker. Please contact the manufacturer if required.
1)Repair or change the socket
2)Replace the indicator
Normal power
voltage, but
the indicator
doesn't
illuminate
1)Loose socket
2)Indicator damaged
By the
specialized
maintenance
worker(Refer
to Electric
Systematic
Diagram)
1)Overflow device
shut-off
2)Tube blockage
3)Air filter blockage
Negative
pressure >
0.04 MPa,
with distinct
reduction or
disappearing
of suction
force at tube
outlet
1)Empty the
holder timely
1)After shut-off, turn the
regulating valve loose
counterclockwise to release
negative pressure in tube,
then re-screw
2)Dredge, clean or replace the
tube
3)Replace it with air filter
produced by us.
Solution
Remark
Probable reasons
Problem
Precautions
Transportation and storage conditions
Enviroment Temperature: -40℃~+55℃
Relative Humidity: 10%~93%
Atmospheric Pressure: 700hPa~1060hPa
Note: It is required to store the aspirator in the well-ventilated room without
corrosive gas, and avoid any violent shock while handling.
Electric repair to be conducted by the specialized operator.
Figure 3: Electric Systematic Diagram
Electric systematic diagram(See Figure 3)