background image

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic 

propagation is affected by absorption and reflection from structures, 

objects, and people.

a)Field strengths from fixed transmitters, such as base stations for radio 

(cellular/cordless) telephones and land mobile radio, AM and FM radio 

broadcast, and TV broadcast cannot be predicted theoretically with 

accuracy. To assess the electromagnetic environment due to fixed RF 

transmitters, an electromagnetic site survey should be considered. If the 

measured field strength in the location in which the 7E-B5 is used

exceeds the applicable RF compliance level above, the 7E-B5 should be 

observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as reorienting or 

relocating the aspirators.

b)Over the frequency range 150 kHz to 80MHz, field strengths should be 

less than 3 V/m.

Table 4 Recommended separation distances

Recommended separation distance between portable and mobile RF 

communications equipment and the 7E-B5

The 7E-B5 is intended for use in an electromagnetic environment in 

which radiated RF disturbances are controlled. The customers or the 

users of these YUWELL aspirators can help prevent electromagnetic 

interference by maintaining a minimum distance between portable and 

mobile RF communications equipment (transmitters) and the 7E-B5 as 

recommended below, according to the maximum output power of the 

communications equipment.

Separation distance according to frequency of

transmitter in meter

150kHz~80MHz
d=1.2

150kHz~80MHz
d=1.2

Output Power of

Transmitter in

Watt

Separation distance according to frequency of

transmitter in meter

800MHz~2.5GHz
d=2.3

0.01                        0.12                        0.12                        0.23

0.1                          0.38                        0.38                        0.73

 1                             1.2                          1.2                           2.3

10                            3.8                          3.8                           7.3

100                          12                           12                             23

For transmitters rated at a maximum output power not listed above, the 

recommended separation distance d in meters (m) can be estimated 

using the equation applicable to the frequency of the transmitter, where 

P is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer.

Note: At 80MHz and 800MHz, the separation distance for the higher 

frequency range applies.

Note: These guidelines may not apply in all situations. Electromagnetic 

propagation is affected by absorption and reflection from structures, 

objects, and people.

All specifications and product configurations are subject to change without
 notification.

15

14

Summary of Contents for 7E-B5

Page 1: ...Phlegm Suction Unit JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Sales Center Address Huanyuan East Road No 1 Xuzhuang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address Yu...

Page 2: ...only describes the usage of the product Do not use accessories other than the manufacturer s recommendation This will reduce the performance of the machine 4 Please do not use the machine and return...

Page 3: ...agram 2 Product Features Installing and Commissioning The customer shall carefully inspect if the appearance of product is good and the varieties quantities of the attachments are in conformity with t...

Page 4: ...he holder will not ascend when the suction tube conductor is placed again into the liquid as suctioned and for Item 2 the liquid level still ascends Thus it is required to observe carefully and lift i...

Page 5: ...in the air filter which leads to gradually darkening of the color of filter diaphragm and obviously reducing or even disappearing of suction force at the inlet of tube while the negative pressure indi...

Page 6: ...ng and application Wipe the case outer surface with lightly wet rag already soaked in the disinfectant and prevent any liquid seeping into the pump Never wipe the places marked with letters and patter...

Page 7: ...Operator due to electromagnetic disturbances Attachments 11 10 1 Suction conductor length 2m 7 12 1 pc 2 Suction catheter F8 F12 1 pc respectively of child adult 3 Power cord 1 pc 4 User s manual 1 pc...

Page 8: ...customer or the user should assure that it is used in such environment Guidance and manufacturer s declaration electromagnetic immunity 3 A m Power frequency magnetic fields should be at levels charac...

Page 9: ...etic environment in which radiated RF disturbances are controlled The customers or the users of these YUWELL aspirators can help prevent electromagnetic interference by maintaining a minimum distance...