2. Cleaning
6. Separate the air hose from the nose mask.
5
6
5. Hold both the PAP device and the heated
humidifier tightly, and press the hook button to
separate the two.
The PAP device should be cleaned at least once a week with the following steps. Please refer to the
relevant details for mask cleaning in the manual.
Note
Do not use a dishwasher or a cleaning machine nor any cleaner containing fragrance or adjusting
agent for cleaning.
Any dust on the exterior of the PAP device and the heated humidifier should be cleaned gently with
clean and wet cloth.
Clean the interior of the humidifier tank and the air hose by filling them with the warm water (the
temperature of which should be less than 41℃).
Rinse the humidifier tank and the air hose completely, and dry then in the air.
Wipe the device with a dry and clean cloth.
Note
Ensure that the size of the air filter cotton is proper, so as to prevent water or dust from going into
the PAP device.
B. Air hose inspection:
Inspect the air hose for any damage and crack. Or replace it with a new one.
C. Air filter inspection:
Inspect the air filter cotton once a week, and it is recommended to replace the air filter once every 4
weeks.
If the air filter is damaged or clogged by particulate dusts, replace it with a new one immediately.
D. Replacement steps of Air filter :
1. Open the filter cover, and remove the old
filter cotton.
2. Place the new air filter cotton on the filter
cover, and then install them onto the PAP
device.
1
2
Air filter cotton
If its separator's seal is damaged or cracked, the seal should be replaced with a new one.
The humidifier tank, air hoses, air filters should be inspected regularly.
A. Humidifier tank inspection:
�
If there is any break or crack in the humidifier tank, the tankit should be replaced with a new one.
3. Inspection