30
31
Static mode
This function is mainly for professional
personnel to enter the static mode to test the
monitor through standard pressure gauge.
Warning
: N
ormal
u
sers
d
on
't n
eed
t
o
k
now
this function and also do not operate. The
company will not take any responsibility for
damage caused by this operation.
Press the" "button
after battery
installation, then the screen will show the " "
icon, which means the system is in restore
testing. Several seconds later, the " " icon
disappears and the air pump starts inflating at
the same time, which indicates the test ended.
Then press the " "button to stop
inflating and take out the batteries to enter
the next step.
System restores
Static mode
Press the " "button and hold, meanwhile
install the batteries. Hold
on for about 3
seconds then release the" " button. Then
screen will show the pressure value " " , the
date and time. Now the system has restored
and entered the static mode. Now can take
the static test.
Entering the static mode
After entering the static mode, if the screen
still doesn't show " ", please operate again as
the System restore. Please contact with the
local distributor if it still does not work.
The monitor will
automatically power
off if there is no
operation in 4
minutes.
Tips!
Note: It must restore the system before
entering the static mode, otherwise it may
cause inaccurate results.
Summary of Contents for YE680B
Page 20: ...The icon twinkle 36 37 300mmHg...