background image

 

 21

Troubleshooting

Thermometer Reset

If you’re having issues getting a temperature reading, you may need to reset 

the thermometer and probe. To do that, remove the batteries from the sensors 

and the display screen at the same time. Put batteries back in the 

display 

screen first

, followed by the probe.

Keep in mind that the display screen may lose signal if it is too far away from 

the base station, or if the unit heats above 140°F. If this happens, simply follow 

the reset instructions above.

Thermometer Probe Channel

If you aren’t able to receive a signal from the thermometer probe you 

may need to set the probe channel. To do that, take the back off of each 

thermometer probe and change the channel by sliding the button to position 

1, 2, or 3. Each probe needs to be on its own channel (1, 2, 3) in order to 

monitor treatment correctly. 

Safety Tip Over Switch

If you plug in the ZappBug thermal units and they do not turn on, first check that 

the 

ZappBug thermal unit top switch is ON. Next, check the safety tip over switch 

which is a small black button on the bottom of the red square ZappBug thermal 

unit. This button needs to be pressed in for the ZappBug thermal unit to turn on and 

is designed so that if the ZappBug thermal unit is knocked over, possibly blocking 

the vents, it turns off for safety. When the ZappBug thermal unit is plugged into an 

appropriate power outlet and is flat on the ground, the button should naturally be 

pressed in and the ZappBug thermal unit should turn on.

Getting Up to Temperature

It can take a while for the unit to fully heat up (sometimes 2-3 hours or more 

depending on ambient temperature or the size of the treatment load).

The most common reasons why the unit struggles to get up to 120°F are 

ambient conditions and item arrangement. The ZappBug Room works best in 

rooms that are 70°F or above. 

Whenever possible do not place the ZappBug Room on concrete flooring. If 

the ZappBug Room must be run on a concrete floor, place insulation such as 

cardboard or area rugs between the ZappBug Room and the floor in order to 

limit heat loss.

Always arrange items as spread out as possible to allow air to flow easily 

throughout the ZappBug Room. If necessary, you may need to purchase 

shelving to more easily elevate small items from the floor and separate them 

from each other.

Circuits Popping

The recirculation fan can be on the same circuit as one o

f the ZappBug thermal 

units, but make sure each ZappBug thermal unit is on its own circuit to avoid 

electrical overload.

Summary of Contents for Room

Page 1: ...ual V3 04 22 KEEP OUT OF REACH OF CHILDREN Manufactured For ZAPPBUG 1700 WESTLAKE AVE N STE 200 SEATTLE WA 98109 EPA Est No 101830 CHN 1 MADE IN CHINA Box 1 of 2 Net Weight 52 lbs 23 5 kg Box 2 of 2 N...

Page 2: ...2014 2022 ZappBug All rights reserved Product specifications subject to change 2 This page is intentionally left blank...

Page 3: ...onnect All Four ZappBug Thermal Units 13 3 Connect Recirculation Fan to Room Frame 13 4 Zip the Room Closed 14 5 Wireless Thermometer Set Up 14 6 Turn on ZappBug Thermal Units 17 7 Insert Items for He...

Page 4: ...4 Important Safety Information You must comply with all safety information and operating instructions when using this product Please read all Safety and Hazard Information outlined in this manual Fail...

Page 5: ...safe use of ANY high power appliances 1 Inspect outlet connection and contact strength Insert a standard two prong plug without ground connection into the outlet Ensure that both prong connections off...

Page 6: ...pBug customer service Additional Safety Information Failure to comply with the safety information presented by this manual may result in severe injury and or damage to property Incorrect assembly or i...

Page 7: ...ROOM Always use provided support blocks and or net to ensure maximum air circulation around your belongings If your belongings are densely packed together it will take significantly more time for the...

Page 8: ...less Thermometer and Three Probes X1 X1 X4 X1 Components Included The following components are required for proper use of this product If you did not receive all of the components listed below contact...

Page 9: ...9 Room Tent Recirculation Fan Accessory Carry Bag Extension Cord X1 X1 X1 X3 Support Block X4...

Page 10: ...2014 2022 ZappBug All rights reserved Product specifications subject to change 10 Use Operation 1 ZappBug Room Assembly Unfold The ZappBug Room with Quilted Side Down Unzip All Five Zipper Pulls...

Page 11: ...11 Use Operation continued Extend Room Frame Legs Unfold All Sides Expose Base Note Extend legs until locking brackets click into position...

Page 12: ...ug All rights reserved Product specifications subject to change 12 Expand Room Frame Use Operation continued Note Expand frame until locking pins click into fixed position on all four legs Assemble Ro...

Page 13: ...ound ZappBug Thermal Unit Observe Do Not Cover markings 2 Connect All Four ZappBug Thermal Units Use Operation continued 3 Connect Recirculation Fan to Room Frame The included recirculation fan mounts...

Page 14: ...ct to change 14 5 Wireless Thermometer Set Up Use Operation continued The ZappBug Room comes with an included thermometer base station and three remote probes Thermometer Base Station Thermometer Prob...

Page 15: ...the different thermometer probe temperatures See Figure 1 They will display in the following sequence CH1 CH2 CH3 Thermometer Base Station Set Up 1 Remove the battery door on the back of the base stat...

Page 16: ...ter Each time it flashes the sensor is transmitting data 2 Close the battery door on the thermometer base station See Figure 2 3 Place the wireless thermometer probe in the deepest most insulated part...

Page 17: ...is on turn the ZappBug thermal unit off by switching the power switch to the OFF position Remove the ZappBug thermal unit from the ventilation tube and allow it to cool for 20 minutes The ZappBug ther...

Page 18: ...locations in items being treated Note If items are densely packed they take longer to reach the desired temperature Better air circulation will heat your items faster and ensure bed bug killing temper...

Page 19: ...complete Once the remote sensor reaches a temperature of 120 F or above run the ZappBug thermal unit for 60 minutes 9 Remove Items After Heat Treatment Hint If your home is infested with bed bugs pla...

Page 20: ...BUGS DEAD TIME IN MINUTES TEMPERATURE INSIDEZAPPBUG LETHALIN1HOUR 1130 F Sources Of ce of the Under Secretary of Defense 2019 Bed Bugs Importance Biology and Control Strategies The Armed Forces Pest M...

Page 21: ...pBug thermal unit to turn on and is designed so that if the ZappBug thermal unit is knocked over possibly blocking the vents it turns off for safety When the ZappBug thermal unit is plugged into an ap...

Page 22: ...ns Please contact ZappBug directly before place of purchase ZappBug s staff are here to help If you have any questions about ZappBug or are experiencing any issues please don t hesitate to contact Cus...

Page 23: ...g any property which is damaged when the ZappBug Room does not work properly 6 The cost of repairing or replacing any property which is damaged due to not following the instructions in the owner s man...

Page 24: ...ZAPP BUG TM ZappBug Customer Service service zappbug com 206 430 5514 We are waiting to assist 9a m 5p m PST Monday Friday...

Reviews: