zebris Medical GmbH
FDM-T Technical Data and Operating Instructions
Page 23/65
3.7
zebris SYNC
The
zebris SYNC
is the standard solution for synchronization of the FDM-T system with
third party measuring devices.
The
SYNC-IN
and
SYNC-OUT
sockets provide i
nput and output for support of „sample by
sample“ In- and Out synchronization. Both sockets provide galvanic protection between
third party systems and FDM-T sensor.
WARNING
Patient’s safety is guaranteed by means of galvanic separation ac-
cording to the provisions of IEC 601-1 when a third party device is
synchronized with the FDM-T system. This allows non-medical
equipment to be synchronized with the FDM-T system as long as
such devices are out of patients reach. Nevertheless the user is
completely responsible for the safety of all third party devices used
in combination with the FDM-T system.
The correct synchronisation of all measurement data has to be veri-
fied in case devices are connected to zebris SYNC which have not
been manufactured by zebris Medical GmbH.
zebris does not accept any liability for correct function and reliability
of the system if the clock signal of external devices does not comply
with the signal specifications provided with in this user manual.
NOTE
The zebris FDM software does not support the SYNC-IN and SYNC-
OUT sockets at time of release of this user manual (12/2012).
For synchronization of the FDM-T system with third party measuring
devices zebris recommends to use the WinFDM-T software (Version
2.5.1 and later).