8.2.2 Vicinity of the patient / test person
In practice a distance of 1.5m surrounding the patient has been approved sufficient for de-
fining the patient's vicinity. Within this distance only medical electrical equipment proved to
be in accordance with DIN EN 60601-1 is allowed to be used.
r = 1,5 m
[1]
[3]
[2]
WARNING
When operating the system, the user [2] must ensure that he does
not touch the PC [3] and the patient [1] at the same time. The same
applies for all other non-medical, electrical components; they may
only be used outside the patient's vicinity.
Furthermore, the user must ensure, never to touch the contacts of
the connectors of the interface box and the patient at the same time.
In case of non-observance, dangerous leakage currents can occur.
The following components of the FDM-T system may be used in the vicinity of the patient:
Treadmill (medical versions)
FDM-T sensor (integrated in the treadmill)
zebris Measuring Systems for medical purposes (e.g. CMS20, DAB Bluetooth)
WARNING
The computer and other non-medical electrical equipment (e.g.
camera equipment, lights) have to be located beyond the reach of
the patients (1.5m).