zebris Medical GmbH
PDM/PDM mobile Technical Data and Operating Instructions
Page 16/42
3.5.2 Synchronization of Video Recordings
The
Sync socket
provides synchronization of platform measurements and videos record-
ed with zebris SYNCCam or SYNCLightCam available as accessories.
For connection of SYNCCam or SYNCLightCam the Video Sync-Control Cable is required:
REF-No
21030316 / Video Sync-Control Cable,
cable length 5
Video-SYNC
Signal Direction: Output (from platform to connected device)
The Video-SYNC signal controls the function of SYNCCam / SYNCLightCam
synchronized with the measurement of the PDM-Platform:
Activates / deactivates illumination of SYNCLightCam
Synchronization of video signal
Electrical Specification
Output Resistance
47
High-Level
≥ 3.0V
Low-Level
≤ 0.8V