System Data
S100 / OPMI pico
Version 7.0
Page 250
G-30-1781-en
Recommended safety distances between portable and mobile
RF communication equipment and the S100 / OPMI pico
The device S100 / OPMI pico is intended for use in an electromagnetic envi-
ronment in which RF disturbances are controlled. The customer or the user of
the device S100 / OPMI pico can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communi-
cation equipment (transmitters) and the S100 / OPMI pico - depending on the
output power of the communication equipment as specified below.
Rated power of the
transmitter [W]
Separation distance depending on transmission frequency [m]
150 KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1st
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed in the above table, the recommended separa-
tion distance d in meters (m) can be determined using the equation indicated for each column, with P being
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac-
turer's specifications.
NOTE 1
NOTE 2
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
influenced by absorption and reflection by structures, objects and persons.
d
1 2 P
,
=
d
1 2 P
,
=
d
2 3 P
,
=
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...