Operation
14-Function Foot Control Panel, Wired or Wireless
Version 6.1
Page 54
G-30-1706-en
Foot control panel operating modes
The foot control panel is equipped with energy management to ensure the
longest possible life cycle of the batteries during wireless operation.
In the event of inactivity, the foot control panel changes to the next-highest
operating mode automatically, following specified duration Operating mode 5
is reached by adopting the rest position. If the FCP is operated or placed in
working position, the FCP switches to operating mode 1.
The table below shows you the various operating modes.
Mode
Conduct on operation
Duration inactive
until transition to
next mode
1
Command is carried out with a delay of
max. 100 ms.
10 s
2
Foot control panel switches to operating
mode 1, command is carried out with a
delay of max. 250 ms.
20 s
3
Foot control panel switches to operating
mode 1, command is carried out with a
delay of max. 500 ms.
2 h
4
Foot control panel switches to operating
mode 1 within 3 s, a function is not
carried out.
30 minutes
5
Foot control panel switches to operating
mode 1 within 5 s, the state indicators
flash briefly, one function is not carried
out.
The establishment of the radio
connection usually takes approx. 5 s,
may take up to 20 s in some cases.