14-Function Foot Control Panel, Wired or Wireless
What to do in the event of malfunctions
Fault
Possible cause
Corrective measure
Reference
Intermittent malfunction
The batteries are empty
Replace old batteries with
new ones.
Use of rechargeable
batteries (accumulators)
Replace with new batteries.
Failure of individual button
functions
Configure button function
differently - this is possible
only if the functions on the
stand are configurable.
Call Service.
--
Failure/fault of the radio link
for the wireless foot control
panel
Establish cable connection if
connection cable present.
Position switch always
detects rest position
Establish cable connection if
connection cable present.
Radio link faulty
Establish cable connection if
connection cable present.
Weak radio signal
Establish cable connection if
connection cable present.
No pairing with FCP
upgrade kit
Perform pairing with the
stand, optionally with the
FCP upgrade kit.