Version 6.1
G-30-1706-en
Page 67
14-Function Foot Control Panel, Wired or Wireless
Care and Maintenance
Disinfection
It may be necessary to disinfect the device surfaces.
The maximum application concentrations are:
–
For alcohol (tested with 2 propanol): 60%
–
For aldehyde (tested with glutaraldehyde): 2%
–
For quaternary compounds (tested with DDAC): 0.2%
Disinfectants with an alcohol concentration higher than 70% can be used to
achieve maximum disinfection.
There is a chance that long-term use of such disinfectants may cause the
device's surfaces to become dull or matte, or may loosen the adhesive labels
on the device without causing them to fall off. However, the use of such
disinfectants will never impair device performance or endanger patients.
NOTE
Damage to the device surfaces!
•
Use an aldehyde and/or alcohol-based disinfectant. The use of
quaternary compounds as additives is acceptable. To avoid damaging the
surfaces, do not use any disinfecting components other than those
specified below.