Version 6.1
G-30-1706-en
Page 53
14-Function Foot Control Panel, Wired or Wireless
Operation
Wired foot control panel checklist
•
Always check the function of the foot control panel before surgery
(without patient!) using the following checklist:
Prerequisites
The stand is switched on.
The FCP upgrade kit (optional) is plugged in.
The connection cable is ready for operation and free from any damage.
The connection cable is connected to the stand, optionally to the FCP
upgrade kit.
The freely configurable buttons are configured.
•
Check all functions of the foot control panel.
Stand and surgical microscope must react properly when controlled via
the foot control panel.
CAUTION
Failure of functions!
For safety reasons, do not use this foot control panel if a function fails.
•
Remedy the cause of the failure (see "What to do in the event of
malfunctions") or contact our service department.
CAUTION
Unintentional operation of the foot control panel!
A height-adjustable OP table may operate buttons on the foot control panel
unintentionally.
•
During set-up and use, ensure that buttons cannot be operated
unintentionally.
NOTE!
Use of drapes
The use of the foot control panel with a drape reduces friction on the floor
and thus limits slip resistance.