14-Function Foot Control Panel, Wired or Wireless
What to do in the event of malfunctions
Version 6.1
G-30-1706-en
Page 63
Wired foot control panel troubleshooting
For your safety
This instrument is a high quality technological product. To ensure optimum
performance and safe working order, we recommend having it checked by
our service representative as part of regular scheduled maintenance.
•
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions", attach a sign to the system
stating it is out of order and contact our service representative.
Malfunctions on the foot control panel
The state indications and the audible information signals help you when
diagnosing the malfunction, see page 24 f.
Fault
Possible cause
Corrective measure
Reference
Foot control panel is
inoperative
Connection cable on the
stand, optionally on the FCP
upgrade kit, not plugged in
Plug connection cable in.
Intermittent malfunction
Failure of individual button
functions
Configure button function
differently - this is possible
only if the functions on the
stand are configurable.
Call Service.
--