ATLAS Model 9000
Appendix C
Confidential and Proprietary
63041A0208
C-5
Service Manual
Step 3: Import and Merge Files into the ATLAS 9000
1.
Start the ATLAS Software by double-clicking the icon on the Windows Desktop. Log in
using the “CTS Operator”.
2.
Run the ATLAS Importer (Tools>Run ATLAS Importer).
3.
Select the
Database
then
Add
Button to initiate the ADD/EDIT ATLAS 900 Series
Database window (Figure 8).
4.
Enter
ATLAS 1
in the Name Field. Click the browse button to the right of location
field to select and open the
Corneal.mdb
file located in the Exams folder on the
Desktop.
5.
Once the fields are populated correctly, click
ADD
.
Figure C.8 Add Edit Database Screen
6.
Click the Start Import Button to start importing the ATLAS 900 series data into the
ATLAS 9000 series instrument database (Figure C.9).
7.
Check the
Skip Archived
box, then click
Start Copying
(Figure C.9).
8.
Once the Process has completed, delete the Exams folder from the ATLAS 9000
Windows Desktop.
Summary of Contents for ATLAS 9000
Page 4: ......
Page 49: ...Diagrams ATLAS Model 9000 Confidential and Proprietary Service Manual 6 2 63041A0208 Notes...
Page 51: ...FIGURE 6 2 ATLAS Model 9000 System Wiring Diagram...
Page 53: ...Parts ATLAS Model 9000 Confidential and Proprietary Service Manual 7 2 63041A0208 Notes...
Page 76: ...Appendix A ATLAS Model 9000 Confidential and Proprietary Service Manual A 4 63041A0208...
Page 112: ...Appendix D ATLAS Model 9000 Confidential and Proprietary Service Manual D 2 63041A0208...
Page 114: ...Appendix E ATLAS Model 9000 Confidential and Proprietary Service Manual E 2 63041A0208...
Page 116: ...Appendix F ATLAS Model 9000 Confidential and Proprietary Service Manual F 2 63041A0208...