3.1.1 Table
3-1
Adjustment / Calibration Requirements Following Replacements
This table identifies checks / adjustments that must be made following removal / replacement
procedures. The numbers listed under
Follow-ups Required
are keyed to the
Follow-up
Actions
listed in the second column. For each
Assembly Replaced
, the follow-up numbers
are listed in the order in which they should be performed, not necessarily in numerical
sequence.
Assembly Replaced Follow-ups
Required
Follow-up Actions
•
Hard Disk Drive ……………...….….1,2,3,20
1 – Calibration File Installation (4.3)
•
Optics Head Assembly ………………….1,20
2 – Acronis Installation (Appendix)
•
All other removals and replacements……..20
3 – System Backup ( Appendix )
4 – Windows Activation (4.4)
20 – System Checkout (2.1)
3.2 Chin Rest Removal
1)
Remove the rear cosmetic cover from the patient support.
2)
Remove the two screws from the base of the chin rest assembly.
3)
Grasp the assembly and lift.
4)
Disconnect the sensor.
Right
side
detail
Left
side
detail
Figure 3.1
Patient
Support
Removal
Remove the two
screws from the
base of the chin
rest assembly.
Summary of Contents for ATLAS 9000
Page 4: ......
Page 49: ...Diagrams ATLAS Model 9000 Confidential and Proprietary Service Manual 6 2 63041A0208 Notes...
Page 51: ...FIGURE 6 2 ATLAS Model 9000 System Wiring Diagram...
Page 53: ...Parts ATLAS Model 9000 Confidential and Proprietary Service Manual 7 2 63041A0208 Notes...
Page 76: ...Appendix A ATLAS Model 9000 Confidential and Proprietary Service Manual A 4 63041A0208...
Page 112: ...Appendix D ATLAS Model 9000 Confidential and Proprietary Service Manual D 2 63041A0208...
Page 114: ...Appendix E ATLAS Model 9000 Confidential and Proprietary Service Manual E 2 63041A0208...
Page 116: ...Appendix F ATLAS Model 9000 Confidential and Proprietary Service Manual F 2 63041A0208...