ATLAS Model 9000
Adjustment and Calibration
Confidential and Proprietary
63041A0208
4-5
Service Manual
Procedure
1)
Switch on the instrument and exit into Windows.
2)
Open up
My Computer > C:\atlas
and locate the file named:
atlas_calibration.xml
(Figure 4.2)
Figure 4.2 Optics Calibration File
3) Replace this file with your new copy of the file. You can either delete the file outright
(move it to the recycle bin) or simply rename it.
Note -
Make sure the replacement file is named ATLAS_calibration.xml or the
ATLAS software will not recognize it. In some cases, the file may be named some
variant of the instrument or camera serial number to help differentiate the
calibration files from each other when not on the instrument.