6.9.2. Changing to OFF mode
Change to OFF mode in case the workstation needs to be reset.
1
Switch off the EHT.
2
Close SmartSEM
®
:
a
Select
File/Exit
from the menu.
A window appears asking for confirmation to close
the session.
b
Confirm by clicking on the
Yes
button.
The electron-optical parameters are filed in a mac-
ro in the individual user directory.
3
Shut down the PC.
4
Press the red
OFF
button.
The red
OFF
button lights.
Computer, electronic components and vacuum system are switched off. The electron optical
column is partially ventilated. A 24 V auxiliary voltage is still present to restart the workstation.