START-UP
Carl Zeiss
Installing standard components
Axio Examiner
38
M60-2-0003 e 05/2012
Adjusting the HAL 100 halogen
3.1.13.2
lamp
(1) Coarse adjustment
•
Loosen the clamping screw (Figure 3-18/
2
), and
remove the ready-to-operate halogen
illuminator from the microscope stand.
•
Switch on the 12 V DC 100 W external power
supply unit (see section 3.2.2).
•
Direct the light beam against a projection
surface (wall) which is at least 3 m away.
Do not look into the illuminator's light
exit opening.
•
Set the adjusting screw (Figure 3-19/
3
) using a
ball head screwdriver (width A/F 3) to reach the
best focus possible for both images of the lamp
filaments on the projection surface.
•
Then, set the adjusting screws (Figure 3-19/
4
and
5
) to a value at which the lamp filaments of
one image exactly fill out the gaps in the
reflector image (Figure 3-19/
1
).
(2) Fine adjustment
•
Attach the microscope lamp and locate it in position using the clamping screw at the microscope
stand again.
•
If required, switch out the transmitted-light diffusing screen, and remove the filter drawer from the
corresponding holder compartment.
•
Focus the
≤
40x lens on the specimen, and find a free object position.
•
Remove the eyepiece, and center the lamp filament and the lamp filament's mirror image in the pupil
image using the adjusting screws (Figure 3-19/
4
and
5
).
•
Set the adjusting screws (Figure 3-19/
3
) to optimize the uniform illumination of the pupil image.
•
If required, switch on the diffusing screen, and insert the filter drawer again.
Figure 3-19 Adjusting the HAL 100 halogen
lamp
Summary of Contents for Axio Examiner
Page 1: ...Operating Manual Axio Examiner Research Microscope...
Page 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Page 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Page 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Page 96: ......