START-UP
Axio Examiner
Installing optional components
Carl Zeiss
M60-2-0003 e 05/2012
53
3.3.7
Installing the Senarmont polarizer
•
Remove the stage carrier (see section 3.1.12).
•
Remove the condenser (see section 3.1.10).
•
Remove the condenser carrier (see section
3.1.9), and store it upside down.
•
Screw the mounting frame (Figure 3-39/
2
) using
four screws (Figure 3-39/
4
) to the lower side of
the condenser carrier (Figure 3-39/
1
).
•
Attach the condenser carrier to the stand (see
section 3.1.9).
•
Slide the Senarmont polarizer (Figure 3-39/
3
)
into the guide of the mounting frame.
•
Attach condenser and stage carrier again.
3.3.8
Replacing the DIC prism in the
condenser
•
To take out the cover plate, turn the condenser
cover plate (Figure 3-40/
4
) using the double
function tool (Figure 3-40/
1
) to unlock the
bayonet socket and allow removal.
•
Unscrew the retainer ring (Figure 3-40/
2
) using
the double function tool (Figure 3-40/
1
).
•
Take out the DIC prism (Figure 3-40/
2
).
•
To install the DIC prism, proceed in reverse
order. Take particular care with the correct
orientation of the DIC prism (cams must mesh
with the holder notches). Check that the
lettering at the knurled ring of the revolving
disk is correct.
Figure 3-39 Installing the SÉNARMONT
polarizer
Figure 3-40 Replacing the DIC prism
Summary of Contents for Axio Examiner
Page 1: ...Operating Manual Axio Examiner Research Microscope...
Page 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Page 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Page 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Page 96: ......