OPERATION
Carl Zeiss
Illumination and contrast procedures
Axio Examiner
62
M60-2-0003 e 05/2012
Axio Examiner.A1/.D1/.Z1 with 45 mm
objective:
Use the 11 mm objective adapter for the
Axio Examiner.A1.
•
Lower the objective into the WORK position.
•
Slide a DIC slider (Figure 4-3/
1
, see lettering)
suitable for the objective into the opening
(Figure 4-3/
3
) provided for this above the
objective in the objective adapter.
•
Make certain that the DIC slider locks exactly
into position.
Axio Examiner.D1/.Z1 with W Plan-
Apochromat 20x/1.0 objective, 75 mm
parfocal length:
•
Insert DIC slider SÉNARMONT 75 W PA (426967-
9000-000 or 426967-9070-000) (Figure 4-4/
1
)
into the opening (Figure 4-4/
3
) provided for it in
the objective mount or turret. Ensure correct
orientation as indicated by the DIC slider
lettering.
•
Put the ball head screwdriver into the recess
(Figure 4-4/
2
) in the DIC slider, and push the
DIC slider (Figure 4-4/
1
) into the slot until it
locks into place.
Figure 4-3 Inserting a DIC slider into an
objective adapter
Figure 4-4 Inserting a DIC slider into an
objective mount or objective
nosepiece