START-UP
Carl Zeiss
Mounting the tube lens turret
Axio Imager
34
430000-7344-001
M70-2-0020 e 06/2009
3.3
Mounting the tube lens turret
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (3-3/
3
) and remove the tube
1
) upward.
•
Unscrew four fastening screws (3-3/
2
), remove
coupler plate (3-3/
4
) upward and store it for
any future use.
•
6
5
)
and screw it down using the four supplied
fastening screws (3-3/
8
).
•
Unscrew the tube lens by hand from the tube
to be used (3-3/
1
). Store it in the storage box.
•
1
) without tube lens with its
dovetail into the mount of the tube lens turret
(3-3/
6
) and tighten clamping screw (3-3/
7
).
3.4
Inserting eyepieces and auxiliary
microscope
•
Remove both dust caps (3-4/
1
and
4
) from the
binocular tube.
•
Remove both eyepieces (3-4/
2
) from their cases
and insert them into the binocular tube as far as
they will go.
•
The auxiliary microscope (3-4/
3
) can be inserted
into one of the eyepiece sockets of the
binocular tube in place of an eyepiece. It is used
to view aperture diaphragm, phase and dark-
field stops and to center phase and dark-field
stops. Focusing on these diaphragms and stops
is possible by means of the adjustable eye lens,
which can then be locked by means of the
clamp screw.
Fig. 3-3
Mounting the tube lens turret
Fig. 3-4
Inserting eyepieces