OPERATION
ZEISS
TFT display touchscreen on the Axio Observer 7, 7 materials
Axio Observer
120
431004-7244-001
12/2016
TL / RL illumination switch
The
Off
and
On
switches open or close the shutter for reflected light (RL) and transmitted light (TL) or
switch the illuminators on and off.
Definite Focus switch
The
Off
and
On
switches switch Definite Focus on and off.
If the system does not have Definite Focus, the Make it visible! Button will appear in place of
the Off and On switch.
Make it visible! button
If the microscope is so poorly adjusted that no image of the specimen is visible, this button can be used
to reset the microscope to a standard state in which the specimen is visible.
The following settings are adapted in this basic state:
Bio / Med stand:
−
transmitted light illuminator is set to medium brightness (3 V)
−
aperture diaphragm is opened
−
TL shutter open, RL shutter closed
−
Motorized condenser is switched to brightfield
−
reflector turret is rotated to the nearest brightfield position
−
Light path switching is set to 100 % vis
MAT (materials) stand:
−
Reflected light illuminator (HAL 100 or LED) set to medium brightness
−
TL shutter closed
−
RL shutter opened
−
Reflector turret in brightfield reflected light position (alternatively contrast module)
−
reflected light luminous-field diaphragm opened
−
reflected light aperture diaphragm opened
−
Optovar turret in position 1: Optovar 1x
−
Sideport in position 1 (100% vis)
−
Baseport in position 2 (100% vis)
The nosepiece position (and, if entered, the objective) will be shown on the left of the status bar, the
halogen illuminator voltage on the right.
Summary of Contents for Axio Observer Series
Page 1: ...Axio Observer Inverted microscope Operating manual...
Page 192: ......